Electrochemiluminescence Immunoassay of P53 from Human Swo-38 Glioma Cell

GH Yan,YH He,D Xing,YD Wei,SC Tan
DOI: https://doi.org/10.1117/12.462550
2002-01-01
Abstract:Electrochemiluminescence (ECL) technique for bimolecular detection offers ultra-high sensitivity and good signal-to- noise ratio. In this work, an ECL reaction and detection system were established. We used a big working electrode and a single photon counter to heighten the detection limit. The results showed that the detection limit of the system is 1/1000 fmol/L free labeled Ru(bpy)3 2+. The concentration dependence of ECL intensity was not linear in the range of low concentration (fmol/L level) free labeled Ru(bpy)3 2+. While in pmol/L level, the ECL intensity was linear. Ru(bpy)3 2+ + TPA (tripropylmine) ECL reaction system was used to indicate P53 protein from human SWO-38 cell line, which was established from grade 1-11 glioma tumor. This method could have potential applications in early-stage tumor diagnosis and environmental inspection.
What problem does this paper attempt to address?