Abstract:The anti-thrombin aptamer can interact with thrombin in high affinity. This interaction results in the change of the fluorescent polarization in the solution of the aptamer labeled with fluorescence. We developed a new method based on the change of the fluorescent polarization resulted from this interaction to detect thrombin. The linear range of the determination is 0.1 similar to 4 IU/mL. The detection limit for thrombin is 0.09 IU/mL. This method has been used to determine thrombin in human whole blood. The results obtained are satisfactory.