Research On The Scanning System Of Fiber-Optic Confocal Microscope

Xl Wang,Zy Chi,S Wu,Wj Chen
DOI: https://doi.org/10.1117/12.445744
2001-01-01
Abstract:Using dual-galvanometer optical scanning head as the reflecting type optic-fiber confocal scanning microscope's planar scanning outfit can achieve the fast scanning to plane image, but there are distortion and vignetting. In the paper, the basic principle of reflecting type fiber-optical confocal scanning microscope is described briefly. The distortion caused by using the common eyepiece as scanning lens, the inherent distortion and vignetting for pre-eyepiece dual galvanometer scanning system have been analyzed in theory, and the according high-speed scanning electronic control system that has the function of distortion correction and image processing has been discussed.
What problem does this paper attempt to address?