Synchronous Enlargement Algorithm for Image Segmentation in Nuclear Medicine

刘志宇,金永杰,李玉兰
DOI: https://doi.org/10.3321/j.issn:1000-0054.2001.04.005
2001-01-01
Abstract:A new method was proposed for the extraction of the leftventricle boundary in a gated blood pool image. This automated method only uses the position characteristics of the left ventricle and right ventricle to find their central regions, then synchronously enlarges the left ventricle and the right ventricle to find the boundary point between them. The synchronous enlargement algorithm does not employ a complicated transform which needs high quality images and enhancement algorithms, and it does not need to find the maximum gradient point between the left ventricle and the right ventricle. This method avoids the most difficult problems for extracting the left ventricle and has a high success rate for gated blood pool images. The study of 25 cases shows that the method can accurately form all the left ventricle boundaries. Furthermore, the method is easy to be extended to 3-D situation.
What problem does this paper attempt to address?