A Novel Application of Using a Commercial Fraunhofer Diffractometer to Size Particles Dispersed in a Solid Matrix

J Gao,C Wu
DOI: https://doi.org/10.1002/1097-4628(20000801)77:5<1165::aid-app25>3.0.co;2-t
IF: 3
2000-01-01
Journal of Applied Polymer Science
Abstract:After a proper modification of a commercial Coulter LS230 particle size analyzer that is based on Fraunhofer principle, we have developed its novel application in directly sizing particles dispersed inside a solid matrix. Using several particle standards, we have demonstrated its validity by comparing the particle size distribu- tions, respectively, obtained in water and directly inside different solid matrixes. In comparison with other conventional sizing methods, such as transmission electron microscopy (TEM) and scanning electron microscopy (SEM), this novel application of Fraunhofer diffraction provides a fast, convenient, and direct way to characterize particles dispersed inside a transparent or translucent solid matrix. It can be envi- sioned that this application can be extended to directly size fillers inside a polymer materials, microdomains inside a polymer blends, and bubbles inside a foam. © 2000
What problem does this paper attempt to address?