M-Aminophenol-h2o2-horseradish Peroxidase Voltammetric Enzyme-linked Immunoassay for Trace Thyroxine in Human Serum

SS Zhang,J Kui,HY Chen,MX Wang
DOI: https://doi.org/10.3321/j.issn:0253-3820.1999.09.001
1999-01-01
Abstract:A voltammetric enzyme-linked immunoassay based on new system of nt-aminophenol (MAP)-H2O2-horseradish peroxidase (HRP) has been developed and used for the detection of MRP and thyroxine (T-4). HRP or labelled HRP catalyzes the oxidation reaction of MAP with H2O2, the product of which produces a sensitive voltammetric peak at potential of -0.46 V ( vs. SCE) in Britton-Robinson (B-R) buffer solution. By using this voltammetric peak, the detection limit to thyroxine (T-4) is 0.5 mu g/L and the linear range 0.5 similar to 320 mu g/L. The detection limit of T-4 by this technique is lower than that by the enzyme-linked immunosorbent spectrophotometric assay (ELISA). Some human serum samples have been analyzed with satisfactory results.