Assessment of Tear Film Stability in Dry Eye with a Newly Developed Keratograph

Jiaxu Hong,Xinghuai Sun,Anji Wei,Xinhan Cui,Yimin Li,Tingting Qian,Wentao Wang,Jianjiang Xu
DOI: https://doi.org/10.1097/ico.0b013e3182714425
IF: 3.152
2013-01-01
Cornea
Abstract:Purpose: To investigate the applicability of a newly developed corneal topographer in assessing tear film stability.Methods: This is a prospective, case-control study. Forty-four Chinese dry eye patients and 41 normal subjects were recruited. Noninvasive tear film break-up time (NI-BUT) was measured using a new method based on a corneal topographer equipped with modified scan software. The reliability of the measurements was determined. Then, the correlations between the NI-BUT and the traditional fluorescein tear film break-up time, Schirmer I test values, and inferior tear meniscus height measurements were determined. The receiver operating characteristic curve technique was used to evaluate the NI-BUT examination in the diagnosis of dry eye.Results: In total, a significant difference between the NI-BUT and the fluorescein tear film break-up time was found (3.2 +/- 2.3 seconds vs. 5.2 +/- 3.4 seconds; P < 0.001). The coefficient of variation and intraclass correlation coefficient values of NI-BUT were 12.8% and 0.93, respectively, for NI-BUT for intraobserver repeatability and 15.4% and 0.88, respectively for interobserver repeatability. The NI-BUT showed a good correlation with other dry eye examinations (all P < 0.05). In addition, the NI-BUT was significantly shorter in dry eye patients (2.0 +/- 0.2 seconds) than in normal subjects (4.3 +/- 0.3 seconds; P < 0.001). When the cutoff value was set at <2.65 seconds, the sensitivity and specificity of the test were 84.1% and 75.6%, respectively.Conclusions: Measurements of NI-BUT obtained with the newly developed corneal topographer may provide a simple, noninvasive screening test for dry eyes with acceptable sensitivity, specificity, and repeatability.
What problem does this paper attempt to address?