The European comparison of absolute gravimeters 2011 (ECAG-2011) in walferdange, Luxembourg: Results and recommendations

olivier francis,henri baumann,tomas volarik,christian rothleitner,gilbert klein,marc seil,nicolas dando,ray tracey,christian ullrich,stefaan castelein,hu hua,wu kang,shen chongyang,xuan songbo,tan hongbo,li zhengyuan,vojtech palinkas,j kostelecký,jaakko makinen,jyri naranen,s merlet,tristan farah,christine guerlin,franck pereira dos santos,nicolas le moigne,cedric champollion,sabrina deville,ludger timmen,reinhard falk,herbert wilmes,domenico iacovone,francesco baccaro,alessandro germak,emanuele biolcati,jan krynski,marcin sekowski,tomasz olszak,a pachuta,jonas agren,andreas engfeldt,rene reudink,pedro inacio,daniel mclaughlin,geoff shannon,marc eckl,tim wilkins,derek van westrum,ryan billson
DOI: https://doi.org/10.1088/0026-1394/50/3/257
2013-01-01
Metrologia
Abstract:We present the results of the third European Comparison of Absolute Gravimeters held in Walferdange, Grand Duchy of Luxembourg, in November 2011. Twenty-two gravimeters from both metrological and non-metrological institutes are compared. For the first time, corrections for the laser beam diffraction and the self-attraction of the gravimeters are implemented. The gravity observations are also corrected for geophysical gravity changes that occurred during the comparison using the observations of a superconducting gravimeter. We show that these corrections improve the degree of equivalence between the gravimeters. We present the results for two different combinations of data. In the first one, we use only the observations from the metrological institutes. In the second solution, we include all the data from both metrological and non-metrological institutes. Those solutions are then compared with the official result of the comparison published previously and based on the observations of the metrological institutes and the gravity differences at the different sites as measured by non-metrological institutes. Overall, the absolute gravity meters agree with one another with a standard deviation of 3.1 mu Gal. Finally, the results of this comparison are linked to previous ones. We conclude with some important recommendations for future comparisons.
What problem does this paper attempt to address?