Amperometric Biosensor for NADH and Ethanol Based on Electroreduced Graphene Oxide–polythionine Nanocomposite Film

Zou Li,Yi Huang,Li Chen,Xiaoli Qin,Zhao Huang,Yaping Zhou,Yue Meng,Jing Li,Siyu Huang,Yan Liu,Wen Wang,Qingji Xie,Shouzhuo Yao
DOI: https://doi.org/10.1016/j.snb.2013.01.072
2013-01-01
Abstract:A nanocomposite film is prepared by cyclic voltammetric deposition of electroreduced graphene oxide (ERGO) and polythionine (PTH) on a glassy carbon (GC) electrode. The electrodeposition of ERGO–PTH film is also investigated via the electrochemical quartz crystal microbalance (EQCM) method for more in-depth understanding of the process. The ERGO–PTH nanocomposite film is applied as transducer for facilitated electrocatalytic oxidation of NADH and for the construction of dehydrogenase-based amperometric biosensor. Under optimum conditions, the amperometric detection of NADH provides a wide linear detection range (0.01–3.9mM), a high sensitivity (143μAmM−1cm−2) and a low limit of detection (LOD=0.1μM, S/N=3). With alcohol dehydrogenase (ADH) as a model, we obtain a linear amperometric response of the ADH-modified ERGO–PTH/GC electrode to ethanol concentration from 0.05 to 1.0mM, a sensitivity of 2.8μAmM−1cm−2, and a LOD of 0.3μM (S/N=3).