Development of a capillary flow microfluidic biosensor with on-chip reagent delivery using water-soluble nanofibers

Shengquan Jin,Minhui Dai,Bang-ce Ye,Sam R. Nugen
DOI: https://doi.org/10.1007/s00542-013-1742-y
2013-01-01
Abstract:Although the potential role of microfluidics in point of care diagnostics is widely acknowledged, the practical limitations to their use still limit deployment. Here, we developed a capillary flow microfluidic with on-chip reagent delivery which combines a lateral flow assay with microfluidic technology. The horseradish peroxidase tagged antibody was electrospun in a water-soluble polyvinylpyrrolidone nanofibers and stored in a microfluidic poly(methyl methacrylate) chip. During the assay, the sample containing on immunomagnetic beads came in contact with the nanofibers causing them to dissolve and release the reagents for binding. Following hybridization, the solution moved by capillary flow toward a detection zone where the analyte was quantified using chemiluminescence. The limit of detection was found to be approximately 10 CFU/mL of . More importantly, the ability to store sensitive reagents within a microfluidic as nanofibers was demonstrated. The fibers showed almost instant hydration and dissemination within the sample solution.