SPR double-differential interferometric imaging system for biomolecular interaction analyses

余兴龙,王大千,张玮,罗昭锋,邓焱,丁丽丽,欧惠超,闫硕,章恩耀
DOI: https://doi.org/10.16511/j.cnki.qhdxxb.2013.02.002
2013-01-01
Abstract:High-throughput and high-sensitivity biomolecular interaction analysis instruments are urgently needed by drug screening and proteomics research. An instrument was designed and implemented based on a novel surface plasmon resonance (SPR) double-differential interferometric imaging method and the concept of refractive index related factor (RIRF). After sequentially passing through a half wave plate, a lens and a polarization beamsplitter, the light reflected from SPR sensing surface splits up into two orthogonal beams and forms two complementary interferograms of the sensing surface at the same time, which are received by two CCD cameras simultaneously. The RIRF distribution of the sensing surface was calculated based on these collected images with the information of biomolecular interaction obtained. Experimental results indicated that the dynamic range of this instrument is 0.015 RIU (refractive index unit) with the sensitivity of 57.15 RIRF/RIU, the resolution of 8.5×10-6 RIU and the throughput of more than 100. The repeatability is better than 95%. The experimental results of the instrument are parallel with those of a Biacore instrument.