Low-background interference detection of glyphosate, glufosinate, and AMPA in foods using UPLC-MS/MS without derivatization
Jun Dong,YiQing Hu,XiaoLu Su,YanXing Yao,Qian Zhou,MengYue Gao
DOI: https://doi.org/10.1007/s00216-024-05158-3
2024-02-29
Analytical and Bioanalytical Chemistry
Abstract:The abuse of herbicides has emerged as a great threat to food security. Herein, a low-background interference detection method based on UPLC-MS was developed for the simultaneous determination of glufosinate, glyphosate, and its metabolite aminomethylphosphonic acid (AMPA) in foods. Initially, this study proposed a simple and rapid pretreatment method, utilizing water extraction and PRiME HLB purification to isolate glyphosate, glufosinate, and AMPA from food samples. After the optimization of pretreatment conditions, the processed samples are then analyzed directly by ultra-performance liquid chromatography–tandem mass spectrometry (UPLC-MS) without pre-column derivatization. The method can effectively reduce interference from by-products of pre-column derivatization and background substrates of food sample, showing low matrix effects (ME) ranging from − 24.83 to 32.10%. Subsequently, the method has been validated by 13 kinds of food samples. The recoveries of the three herbicides in the food samples range from 84.2 to 115.6%. The limit of detection (LOD) is lower to 0.073 mg/kg, 0.017 mg/kg, and 0.037 mg/kg, respectively. Moreover, the method shows an excellent reproducibility with relative standard deviations (RSD) within 16.9%. Thus, the method can provide high trueness, reproducibility, sensitivity, and interference-free detection to ensure human health safety.
chemistry, analytical,biochemical research methods