An Intravascular Loopless Monopole Antenna for Vessel Wall MR Imaging at 3.0 T

Hongyang Yuan,Xing Lv,Xiaohai Ma,Rui Zhang,Youyi Fu,Xuedong Yang,Xiaoying Wang,Zhaoqi Zhang,Jue Zhang,Jing Fang
DOI: https://doi.org/10.1016/j.mri.2012.06.032
IF: 3.13
2013-01-01
Magnetic Resonance Imaging
Abstract:The purpose of this study was to develop a novel intravascular loopless monopole antenna (ILMA) design specifically for imaging of small vessel walls. The ILMA consisted of an unshielded, low-friction guide wire and a tuning/matching box. The material of the guide wire was nitinol and it was coated with polyurethane. Because the guide wire was unshielded, it could be made thinner than the coaxial cable-based loopless intravascular antenna design. The material of the box was aluminum. In this study, the diameter of the guide wire was 0.5 mm and the length was 58.7 mm. The ILMA was used as a receiving antenna and body coil for transmission. To verify the feasibility of the ILMA, in vitro and in vivo experiments were performed on a 3.0-T magnetic resonance (MR) scanner. In vitro tests using the ILMA indicated that the proposed design could be used to image target vessel walls with a spatial resolution of 313 μm at the frequency coding direction and more than 100 mm of longitudinal coverage. In vivo tests demonstrated that the images showed the vessel walls clearly by using the ILMA and also indicated that the ILMA could be used for small vessels. The proposed antenna may therefore be utilized to promote MR-based diagnoses and therapeutic solutions for cardiovascular atherosclerotic diseases.