Depth-recognizable time-domain fluorescence molecular tomography in reflective geometry
Jiaju Cheng,Peng Zhang,Chuangjian Cai,Yang Gao,Jie Liu,Hui Hui,Jie Tian,Jianwen Luo
DOI: https://doi.org/10.1364/boe.430235
IF: 3.4
2021-06-04
Biomedical Optics Express
Abstract:Conventional fluorescence molecular tomography (FMT) reconstruction requires photons penetrating the whole object, which limits its applications to small animals. However, by utilizing reflective photons, fluorescence distribution near the surface could be reconstructed regardless of the object size, which may extend the applications of FMT to surgical navigation and so on. Therefore, time-domain reflective fluorescence molecular tomography (TD-rFMT) is proposed in this paper. The system excites and detects the emission light from the same angle within a field of view of 5 cm. Because the detected intensities of targets depend strongly on the depth, the reconstruction of targets in deep regions would be evidently affected. Therefore, a fluorescence yield reconstruction method with depth regularization and a weighted separation reconstruction strategy for lifetime are developed to enhance the performance for deep targets. Through simulations and phantom experiments, TD-rFMT is proved capable of reconstructing fluorescence distribution within a 2.5-cm depth with accurate reconstructed yield, lifetime, and target position(s).
radiology, nuclear medicine & medical imaging,biochemical research methods,optics