Quantitative Determination of Rivastigmine in Rat Plasma and Tissues by HPLC-MS/MS Method

张彦青,仰浈臻,王占璋,吴凯,齐宪荣
2012-01-01
Abstract:OBJECTIVE: To establish a highly accurate, sensitive and credible high performance liquid chromatography-tandem mass spectrometry (HPLC-MS/MS) method for the determination of rivastigmine in rat plasma and tissues. METHODS: All biological samples were extracted with ethyl acetate. Chromatographic separation was performed on an Agilent C 18 column (4.6 mm × 100 mm, 3.5 μm) using water-methanol (20:80) as mobile phase at a flow rate of 0.3 mL · min -1. The MS analysis was performed by multiple reaction monitoring (MRM) with electronic spray ionization source (ESI +) for quantitative response of rivastigmine (251.3 → 206) and the internal standard, antipyrine (189.1 → 106.1) respectively. RESULTS: The calibration curves of rivastigmine in plasma and tissues showed satisfactory linearity, and the relative extraction recovery of three concentrations in plasma and all tissues were over 60% except some individuals and all RSDs met the requirements. The limit of quantification of rivastigmine in plasma and tissues were 0.1 and 0.2 ng · mL -1 respectively. CONCLUSION: The method is rapid, accurate, highly selective and sensitive, and is suitable for the quantitative analysis of rivastigmine in biological sample of rats. Copyright 2012 by the Chinese Pharmaceutical Association.
What problem does this paper attempt to address?