Discrepant Findings of Computed Tomography Quantification of Minimal Lumen Area of Coronary Artery Stenosis: Correlation with Intravascular Ultrasound

Yuehua Li,Jiayin Zhang,Zhigang Lu,Jingwei Pan
DOI: https://doi.org/10.1016/j.ejrad.2012.03.010
IF: 4.531
2012-01-01
European Journal of Radiology
Abstract:OBJECTIVES:To study the diagnostic performance of computed tomography (CT) quantification of minimal lumen area (MLA) based on multiple factors (image quality, calcification and lesion locations), with reference to intravascular ultrasound (IVUS).METHODS:Consecutive clinically ordered patients were prospectively enrolled in our study. CT quantification of MLA was manually measured on cross-sectional view and further compared with IVUS findings. A significant lesion was defined as ≤6 mm2 MLA for the left main (LM) coronary artery and a ≤4 mm2 MLA for other epicardial vessels.RESULTS:Non-calcified lesions had good correlation between CT and IVUS (r=0.96) and the sensitivity, specificity, positive predictive value (PPV) and negative predictive value (NPV) were 85% (51/60), 93.94% (31/33), 96.23% (51/53) and 77.5% (31/40) respectively. Decreased correlation (r=0.814) lower specificity (23.81% (5/21)) and positive predictive value (60.98% (25/41)) were observed in calcified subgroup. Artifact-absent lesions demonstrated excellent correlation (r=0.967) and the sensitivity, specificity, PPV and NPV were 96% (48/50), 93.94% (31/33), 96% (48/50) and 93.94% (31/33) respectively. Impaired correlation (r=0.584) was noted in artifact-present lesions and the sensitivity, specificity, PPV and NPV were 80% (28/35), 23.81% (5/21), 63.64% (28/44) and 41.67% (5/12) respectively. Excellent correlation between CT and IVUS was noted in proximal epicardial vessels (r=0.908) and the sensitivity, specificity, PPV and NPV were 92.31% (36/39), 78.13% (25/32), 83.72% (36/43) and 89.29% (25/28) respectively. Middle right coronary artery lesions showed non-significant correlation (r=0.54, p=0.055).CONCLUSIONS:CT quantification of MLA can only be accurately achieved in non-calcified lesions with absence of artifact. Diagnostic performance is impaired in calcified lesions.
What problem does this paper attempt to address?