Fast and High-Performance Screening of Narcotic Drugs on A Microfluidic Device by Micellar Electrokinetic Capillary Chromatography

Jin Sheng,Qu Ping,Jianping Lei,Huangxian Ju,Chaojin Song,Daming Zhang
DOI: https://doi.org/10.1080/00032719.2011.653894
2012-01-01
Analytical Letters
Abstract:This work presents a new approach for fast and sensitive ultraviolet detection of 12 kinds of narcotic drugs on a microfluidic device by micellar electrokinetic capillary chromatography. Under optimal sampling and separation conditions the baseline separation of 12 drugs with resolution values ranging from 1.06 to 4.04 and separation efficiency up to 5.14 x 10(5) plates m(-1) was achieved within 200 s. The widest linear range for detection of these analytes was 1.0 to 1500.0 mu g mL(-1). The correlation coefficients are higher than 0.9994. This system can successfully be applied to analyze narcotic drugs in human urine with the aid of liquid-liquid extraction of the samples. This method allows minimum detectable concentrations of these drugs down to 33 ng mL(-1) at a signal-to-noise ratio of 3. This rapid method with high resolution and sensitivity, and little solvent consumption possessed potential application in screening of narcotic drugs in forensic analysis.
What problem does this paper attempt to address?