A Highly Sensitive and Selective Method for Dopamine Detection Based on Poly (folic Acid) Film Modified Electrode

Maofang He,Xiaohui Zheng
DOI: https://doi.org/10.1016/j.molliq.2012.05.024
IF: 6
2012-01-01
Journal of Molecular Liquids
Abstract:A highly sensitive and selective method based on poly (folic acid) modified electrode (PFA/CPE) to detect dopamine (DA) was established by electrochemical polymerization of folic acid on a carbon paste electrode (CPE). Subsequently, the morphologies of the electrode surface were observed by scanning electron microscopy (SEM) and the electrochemical behavior of DA at the PFA/CPE was studied. It was illustrated that the PFA/CPE had excellent electrocatalytic ability towards the oxidation of DA. The anodic peak current of DA was greatly enhanced at the PFA/CPE and the standard rate constant (ks) could be calculated to be 3.35s−1. Moreover, the interferences of ascorbic acid (AA) and uric acid (UA) on the determination of DA were eliminated. Consequently, the PFA/CPE was subjected to the determination of DA in the presence of AA and UA successfully. The linear responses were obtained over two concentration internals: 2.5×10−8–1.0×10−6molL−1 and 1.0×10−6–1.0×10−3molL−1 with a detection limit (S/N=3) of 8×10−9molL−1. With favorable selectivity and sensitivity, the present method has been applied to the determination of DA in dopamine hydrochloride injections.
What problem does this paper attempt to address?