A Flow Injection Fluorescence "Turn-On" Sensor for the Determination of Metformin Hydrochloride Based on the Inner Filter Effect of Nitrogen-Doped Carbon Dots/gold Nanoparticles Double-Probe.
Guo-Qi Zhang,Xu-Yan Zhang,Ya-Xiong Luo,Yong-Sheng Li,Yang Zhao,Xiu-Feng Gao
DOI: https://doi.org/10.1016/j.saa.2020.119384
IF: 4.831
2020-01-01
Spectrochimica Acta Part A Molecular and Biomolecular Spectroscopy
Abstract:In this paper, an ultrasensitive and rapid "turn-on" fluorescence sensor, integrating flow-injection (FI) with nitrogen-doped carbon dots/gold nanoparticles (N-CDs/AuNPs) double-probe is established for the determination of metformin hydrochloride (MET) in biological fluids. The sensing strategy involves the weak inner filter effect between AuNPs and N-CDs due to aggregation products of MET with AuNPs. Unfortunately, the degree of AuNPs aggregation is difficult to control through manual assays, resulting in intolerable measurement error that limits further applications. However, the proposed method overcomes the above problem, and significantly lowers the consumption of expensive reagents (AuNPs: about 60 lL per test). Under optimal conditions, the fluorescence intensity at 400 nm excitation and 505 nm emission wavelengths display a linear correlation with MET concentration (5-100 mu g L-1) and the limit of detection is 2.32 mu g L-1 (3.3 S/k). The advantages of the presented method include high sensitivity, rapid speed (60 sample h(-1)), good accuracy and precision (RSD <= 2.1%, n = 11) and low cost. Since MET is the first-line hypoglycemic agent in patients with type II diabetes, this method can preliminarily determine MET content in urine samples, giving satisfactory results. (C) 2020 Elsevier B.V. All rights reserved.