Multi-Parameter Detection of Urine Based on Electropolymerized PANI: PSS/AuNPs/SPCE
Dong Wang,Xiyu Mao,Yitao Liang,Yu Cai,Tingting Tu,Shanshan Zhang,Tianyu Li,Lu Fang,Yue Zhou,Zhaoyang Wang,Yu Jiang,Xuesong Ye,Bo Liang
DOI: https://doi.org/10.3390/bios13020272
2023-01-01
Biosensors
Abstract:Urine analysis is widely used in clinical practice to indicate human heathy status and is important for diagnosing chronic kidney disease (CKD). Ammonium ions (NH4+), urea, and creatinine metabolites are main clinical indicators in urine analysis of CKD patients. In this paper, NH4+ selective electrodes were prepared using electropolymerized polyaniline-polystyrene sulfonate (PANI: PSS), and urea- and creatinine-sensing electrodes were prepared by modifying urease and creatinine deiminase, respectively. First, PANI: PSS was modified on the surface of an AuNPs-modified screen-printed electrode, as a NH4+-sensitive film. The experimental results showed that the detection range of the NH4+ selective electrode was 0.5~40 mM, and the sensitivity reached 192.6 mA M-1 cm-2 with good selectivity, consistency, and stability. Based on the NH4+-sensitive film, urease and creatinine deaminase were modified by enzyme immobilization technology to achieve urea and creatinine detection, respectively. Finally, we further integrated NH4+, urea, and creatinine electrodes into a paper-based device and tested real human urine samples. In summary, this multi-parameter urine testing device offers the potential for point-of-care testing of urine and benefits the efficient chronic kidney disease management.