Measurement Accuracy of Temporomandibular Joint Space in Promax 3-Dimensional Cone-Beam Computerized Tomography Images.

Zhi-ling Zhang,Jun-ge Cheng,Gang Li,Ji-zong Zhang,Zu-yan Zhang,Xu-Chen Ma
DOI: https://doi.org/10.1016/j.oooo.2011.11.020
IF: 2.538
2012-01-01
Oral Surgery Oral Medicine Oral Pathology and Oral Radiology
Abstract:Objective. The aim of this study was to evaluate and compare the measurement accuracy of the temporomandibular joint (TMJ) space in the Promax 3D cone-beam computerized tomography (CBCT) images scanned with 2 different dental protocols.Study Design. TMJ space impression models were made according to the occlusion. Forty joints were scanned with the standard and the large view protocol of the Promax 3D CBCT scanner. Two observers measured the joint spaces 3 times on both radiographs and the photocopies of the impression models.Results. A total of 120 CBCT images were measured. There were no significant differences among the actual joint spaces and the CBCT measurements performed with the 2 scanning protocols (P = .305). The inter- and intraobserver variabilities were not significant.Conclusions. The 2 scanning protocols provided by the Promax 3D CBCT scanner were reliable and similar for recording the TMJ space. (Oral Surg Oral Med Oral Pathol Oral Radiol 2012;114:112-117)
What problem does this paper attempt to address?