A Study on SPR-Based Gene-Chip System for Rapid Detection of Neisseria Gonorrhoeae

顾大勇,石磊,禹华伟,梁冰,鲁卫平,周元国,徐云庆,史蕾,张雅鸥
DOI: https://doi.org/10.3969/j.issn.1006-7159.2007.03.003
2007-01-01
Abstract:Objective Application of SPR-based gene-chip system for rapid detection of Neisseria gonorrhoeae.Methods The SPR-based gene chip of Neisseria gonorrhoeae was prepared and the optimal fixing time and concentration of probes,the sensitivity and the specificity of the chip were verified.Finally, the SPR-based gene chip was utilized to detect clinical samples.Results When fixed by probes for 4.5 h at concentration of 1500nM,the gene chip shows a well detection performance.It can identify accurately the mismatch between bases and its detection sensitivity is 10fmol/L.When compared with the conventional analysis,the detection results of the chip system were essentially the same.Conclusion The detection results proved that this chip system is accurate and reliable.The chip will have a promising prospect in application.
What problem does this paper attempt to address?