Measurement of Pulmonary Flow Velocity: a Comparison of Doppler Ultrasound and MRI

LEI Xiao-yan,SHEN Ya-mei,GUO You-min,CHEN Min,YANG Jian,XU Gui-ping
DOI: https://doi.org/10.3969/j.issn.1002-1671.2006.01.008
2006-01-01
Abstract:Objective To investigate the accuracy and applied value of measurement of pulmonary flow velocity with phase-contrast MRI(PC-MRI).Methods ①42 healthy volunteers were divided into 2 groups randomly,one group was examined by PC-MRI,another was examined by doppler ultrasound.The inner diameter,maximum velocity and mean velocity of main pulmonary artery(MPA) were measured;②The flow volume of MPA in a cardiac cycle was calculated with formulation.Results ①The inner diameters of MPA measured with PC-MRI and doppler ultrasound were(24.4±3.41) mm and(21.5±2.55) mm respectively;②The maximum velocity and mean velocity of MPA measured by PC MRI were(79.9±24.0) cm/s and(50.3±7.71) cm/s respectively;③The maximum velocity and mean velocity of MPA gained by doppler ultrasound was(88.8±8.33) cm/s and(53.7± 5.04) cm/s resepectively;④The flow volume of MPA in a cardiac cycle measured by PC MRI and doppler ultrasound was(73.5±6.60) ml and(69.0±10.6) ml respectively.The results were analyzed by group sample t-test,and the level of test was 0.05.Conclusion The blood flow velocity and volume of pulmonary artery can be measured accurately by PC-MRI,it is of certain applied value in assessing the pulmonary blood flow dynamics changes.