Highly salicylate-selective membrane electrode based on a new thiomacrocyclic Schiff base complex of binuclear copper(II) as neutral carrier

En-Ping Luo,Ya-Qing Chai,Ruo Yuan,Jian-Yuan Dai,Lan Xu
DOI: https://doi.org/10.1016/j.desal.2008.12.054
IF: 11.211
2009-01-01
Desalination
Abstract:New binuclear metallic complexes of thiomacrocyclic Schiff base, 3,9,13,19 -tetraphenyl-6,16-dithione-1,11-dithio-4,5,7,8,14,15,17,18-octaazacycloeicosa-3,8,13,18-tetraene binuclear metal(II) [M(II)2-TDDOCT] (M=Cu, Co, Ni), were synthesized and their anion response characteristics were investigated. The performances of the electrodes are considerably influenced by the nature of the central metals. The Cu(II) complex-based electrode exhibited a good selectivity to salicylate anion with an anti-Hofmeister selectivity pattern: Sal−>ClO4−>SCN−>I−>Benzoate>Br−>Acetate>F−>SO32−>NO2−>Cl−>NO3−>SO42−>H2PO4−. The electrode had an excellent linear response to Sal− from 9.0×10−7 to 1.0×10−1M with a slope of −59.3mV per decade,a detection limit of 5.0×10−7M, and a fast response time within 15s over the entire concentration series in phosphate buffer solutions of pH 5.0 at 25°C. Spectroscopic techniques and the influence of lipophilic charged additives on the electrode behavior were used to investigate the response mechanism to Sal−. The electrode can be applied to the direct determination of salicylate in human urine and pharmaceutical samples and the results obtained are in accord with the results from a standard method.