Numerical Investigation of Effects of Drug-eluting Stent Link on Flow Field and Concentration Field of Hemoynamics

CHEN Xi,JIANG Wen-tao,FAN Yu-bo,CHEN Yu,ZHENG Ting-hui,GUO Meng,CHEN Jun-kai
2010-01-01
Abstract:Objective To study the effects of drug-eluting stent(DES) links on hemodynamics for better design of the DES. Methods Two DESs grafted into the coronary artery, one with 3 links and another with 6 links, were taken to be numerically simulated and compared about their mechanical performance. The wall shear stress(WSS) and the drug concentration of models were investigated in detail. Results The low WSS region of 6 links model(model B) was 2.62% larger than that of 3 links model(model A) and its low drug concentration region was 4.27% smaller. As a drug carrier more even distribution of structure model B was superior to model A. In addition, the area of high risk of restenosis, low WSS and low drug concentration of what is named"double low" area in model B was significantly less than that in model A. Conclusion The study shows that links contribute to drug depositing in the vicinity of the blood vessel wall, and inhibiting in-stent intimal hyperplasia, so it is concluded that 6 links model is better than 3 links model.
What problem does this paper attempt to address?