Direct Electrochemistry and Electrocatalytic Behavior of Horseradish Peroxidase on Attapulgite Clay Modified Electrode
Huihui Chen,Zhe Zhang,Dongqing Cai,Shengyi Zhang,Bailin Zhang,Jilin Tang,Zhengyan Wu
DOI: https://doi.org/10.2116/analsci.27.613
IF: 1.967
2011-01-01
Analytical Sciences
Abstract:A novel third-generation hydrogen peroxide (H 2 O 2 ) biosensor was developed by immobilizing horseradish peroxidase (HRP) on a biocompatible attapulgite (ATP) modified glassy carbon (GC) electrode. The ATP could provide a biocompatible microenvironment for enzyme molecules, greatly amplify the coverage of HRP molecules on the electrode surface, and most importantly facilitate the direct electron transfer between HRP and the electrode. The biosensor construction process was followed by atomic force microscopy (AFM). Cyclic voltammetry was employed to characterize the properties of the biosensor. A linear calibration plot of the enzyme electrode was obtained over the range of 5 µM to 0.3 mM for H 2 O 2 with a detection limit of 5 µM. Furthermore, the biosensor showed high sensitivity, good reproducibility, and fine long-term stability.