Evaluation of Bispectral Index in Assessing the Depth of Sedation in Patients Treated with Mechanical Ventilation

黄絮,李刚
DOI: https://doi.org/10.3969/j.issn.1001-0025.2010.03.005
2010-01-01
Abstract:Objective:To evaluate the correlation between the clinical assessment of sedation using the Ramsay score and sedation-agitation scale(SAS)and the assessment using bispectral index(BIS)in ventilated and sedated ICU patients.Methods:ICU patients requiring mechanical ventilation and sedation were monitored using the Ramsay score,SAS and BIS.The same doctor recorded each BIS at the time of Ramsay score and SAS assessment every half an hour.Results:Data were collected on 20 subjects generating 358 paired readings.Positive correlation between BIS and Ramsay score values was shown with a Spearman Rank coefficient r value of 0.753(P0.05).Similar correlation was observed between BIS and SAS values with r value of 0.771(P0.05). Wide ranges of BIS scores were observed in each level of sedated patients.Conclusion:BIS monitoring may have some role in sedation assessment in ICU.BIS values should be interpreted with caution,further research is necessary to determine the role of BIS monitoring in ICU practice.
What problem does this paper attempt to address?