The New Taper Geometries Design of Fiber Optical Evanescent Wave Biosensor Probe

LIU Qian-qian,XU Xue-mei,LI Wei,YE Jun-hua,WU Ai-jun,WU Jian-hao,WANG Hua
DOI: https://doi.org/10.3969/j.issn.1005-488x.2008.01.005
2008-01-01
Abstract:In order to enhance the detection efficiency of fiber optical evanescent wave biosensor,the excite light's transmission track in the probe of linear taper is analyzed based on the mode matching radius of fiber probe obtained according to the mode matching theory of fiber in given conditions.Combined with the relationship between evanescent wave penetration depth and incidence angle,a new parabola taper fiber probe is put forward.The calculation shows that to maximum the energy excited at uniform sensitive part,incident angle must be minimal at the end of taper and equal to critical angle.By modifying the parameter of parabola taper,incident angle is closer to the critical angle at the taper transition phase when it's still larger than the critical angle,which excites lager penetration depth.The experiment shows that the sensitivity of parabola taper is better than others,which indicates parabola taper can ideally collect and excite fluorescence,and theory matches with experiment well.
What problem does this paper attempt to address?