Rapid Determination of Promethazine Hydrochloride in Medicine by Microfluidic Chip with Contactless Conductivity Detection

周勰,张吟眉,朱嫦琳,陈缵光
DOI: https://doi.org/10.3969/j.issn.1004-4957.2010.09.019
2010-01-01
Abstract:A rapid and simple method was established for the determination of promethazine hydrochloride in injection and tablet by microfluidic chip with contactless conductivity detection.The experimental factors,including type and concentration of buffer and additive,separation voltage and sample injection time were optimized.In the optimal conditions,using 20 mmol·L-1 MES-20 mmol·L-1 His as buffer solution,separation voltage of 2.2 kV and injection time of 10 s,promethazine hydrochloride in the samples was separated and detected in 2 min.The linear relationship for promethazine hydrochloride ranged from 100 mg·L-1 to 1 000 mg·L-1.The linear equation was Y=-502+4.99×103 X with a correlative coefficient of 0.995 6 and a detection limit(S/N=3) for 10 mg·L-1.The relative standard deviation(RSD,n=5) for 0.1 g·L-1 promethazine hydrochloride was 2.1%.The proposed method was successfully applied in the detection of promethazine hydrochloride in three samples.The spiked recoveries of promethazine hydrochloride from real samples at three concentration levels of 100,300 and 500 mg/L were in the range of 96%-101% with RSDs of 1.7%-3.0%.The method could be used in the quality control of promethazine hydrochloride drug with its simplicity,sensitivity and speediness.
What problem does this paper attempt to address?