B-ultrasound Image Registration of HIFU Monitoring Based on Ultrasonic Speckle

LIU Bing,ZHAO Chunliang,WANG Hua,OUYANG Ting,ZHONG Shanshan,CHEN Wei
2010-01-01
Abstract:The existence of phantoms and other ultrasound noises in Ultrasound-Guided HIFU systems' B-mode ultrasound images will hamper the B-mode ultrasound image registration and the locating of the target.In order to solve this problem,this paper proposes a new method of B-mode ultrasound image registration based on the phase correlation of ultrasonic speckle.Firstly,under the principles and the pseudorandom feature of the speckle,the speckle images and registration,the B-mode ultrasound images are picked up by using the phase correlation of ultrasonic speckle through the moving values of images,on the platform of Maltab 7.0.Secondly,the direct registration of the phase correlation is compared with the registration of the speckle in B-mode ultrasound images.Thirdly,the results of registration are analyzed and the measurement is made of the similarity degree.It is shown that the phase correlation of the speckle could achieve the image registration among the B-mode ultrasound images with phantoms and noises.This registration method enjoys a high robustness and accuracy,and is better than the direct registration of phase correlation.The feasibility of the registration based on speckle is demonstrated by experiments.
What problem does this paper attempt to address?