Performance Evaluation Of Self-Made Microdialysis Probes

Xin Liang,Shao Qing,Wang Qiao,Cheng Yi-Yu,Qu Hai-Bin
DOI: https://doi.org/10.3321/j.issn:0253-3820.2007.07.035
IF: 1.193
2007-01-01
Chinese Journal of Analytical Chemistry
Abstract:With peoniflorin used as a model compound, in vitro microdialysis experiments were carried out to evaluate the influence of a series of experimental conditions (perfusate flow rate, concentration of peoniflorin, membrane and method used to determine recovery) on the extraction efficiency of self-made concentric probes. The results revealed that extraction efficiency was different between probes. An exponential decrease in extraction efficiency parallel to the increase in perfusate flow rate was found. Extraction efficiency became concentration-dependent when the flow rate was 8. 0 mu L/min. Statistically significant differences was found between extraction efficiency determined by direct dialysis and retrodialysis. After the self-made probe was applied to continuously sampling in vivo for 6 hours, the permeability of the membrane on the probe did not significantly. The results show that the self-made microdialysis probes can be used for microdialysis in vivo.
What problem does this paper attempt to address?