Development of Photoelectrical Sensor for the Determination of Cadmium

Guoyan Liu
2011-01-01
Abstract:A test strip for the detection of cadmium was developed by immobilizing 1-(4-nitrophenyl)-3-(3-methyl-pyridyl) triazene,a chromogenic agent specific for cadmium,onto a supporting pad.After the test strip was inserted into a self-made photoelectrical sensor,the reflectivity was recorded as the sensor signal.A standard calibration curve was established between reflectivity and cadmium concentration in standard solution.The results showed that this photoelectrical sensor could detect cadmium over the concentration range of 0.0179-0.2857 mmol/L within 2 min and revealed a detect limit of 0.00348 mmol/L.This technique exhibited the advantages of short assay time and convenient operation.Therefore,this photoelectrical sensor has the potential to apply in the on-site detection of cadmium.
What problem does this paper attempt to address?