Enhanced detection of patent foramen ovale by systematic transthoracic saline contrast echocardiography.

Yat-Yin Lam,Cheuk-Man Yu,Qing Zhang,Bryan P Yan,Gabriel Wai-Kwok Yip
DOI: https://doi.org/10.1016/j.ijcard.2010.06.018
IF: 4.039
2011-01-01
International Journal of Cardiology
Abstract:We aimed to evaluate the effectiveness of transthoracic saline contrast echocardiography (TSCE) in detecting patent foramen ovale (PFO).Transesophageal echocardiography (TEE) is semi-invasive and not ideal for PFO screening.112 patients (48 males, 46 ± 14 years) with suspected PFO received intravenous agitated-saline contrast at rest and stress (strain and release phases of Valsalva maneuver and coughing). The presence of interatrial shunting was defined as >5 bubbles appearing in the left heart within 3 cardiac cycles. The stage of the maneuver at which interatrial shunting occurred was recorded. The TSCE findings were validated by TEE.TEE identified PFO in 45% of patients. The sensitivities of TSCE in detecting PFO at rest, during strain and release of Valsalva maneuver, and coughing were 12.0%, 38.0%, 80.0% and 94.0% respectively (each p<0.05 when compared to previous stage). Specificities were similar and >95% for all stages. Moreover, the release phase of the maneuver improved the diagnostic accuracy [defined as (number of true positives+true negatives) divided by total in sample] with incremental value over the preceding strain phase (89.2 vs. 70.5%, p<0.001).Patent foramen ovale can be identified confidently with proper conduct of the Valsalva maneuver during the transthoracic saline contrast echocardiography.
What problem does this paper attempt to address?