Coupling A Microchip with Electrospray Ionization Quadrupole Time-Of-Flight Mass Spectrometer for Peptide Separation and Identification

Hai-Fang Li,Jiangjiang Liu,Zongwei Cai,Jin-Ming Lin
DOI: https://doi.org/10.1002/elps.200700477
2008-01-01
Electrophoresis
Abstract:The present study reports a simple method of coupling a glass microchip to an electrospray ionization (ESI) quadrupole time-of-flight mass spectrometer (QTOF-MS) for separation and identification of peptides. A sheath-flow electrospray interface was constructed based on attaching a short fused-silica capillary to the microchip. The dead volume at the interface was effectively reduced by wet etching an approximate flat-bottom capillary insertion channel coaxial to the end of separation microchannel and using a wire-controlled epoxy-blocking attachment method. The makeup liquid and neb gas were coaxially pumped through two stainless-steel tees to maintain a stable and efficient electrospray. The coupled microchip/ESI-QTOF-MS system was successfully used to carry out electrophoresis separation of peptides and ESI-QTOF-MS identification.