Estimation of velocities of tissue motion and blood flow based on radio frequency signal deconvolution

CHEN Gong,YU Jie,SONG Fu-xian,ZHANG Dong
DOI: https://doi.org/10.3969/j.issn.1000-3630.2010.02.011
2010-01-01
Abstract:A novel approach based on radio frequency signal deconvolution is developed to estimating the velocities of blood flow and tissue motion without MTI filtering.In this method,the sampled sequences of the radio frequency echoes are transformed to the frequency domain by FFT algorithm firstly.Then these transformed sequences are deconvolved.Finally the velocities of the blood flow and the tissue motion can be obtained by a spectrum analysis of these deconvolved sequences.This approach has been applied to the simulated ultrasound echoes with two typical time-delay parameter sets.The validity of this method is examined by the comparison between the estimated velocities of blood and tissue motion and the given values.This approach is helpful to the measurement of the low-velocity blood flows.
What problem does this paper attempt to address?