Determination of Anthracene and Pyrene in Non-Aqueous System by Fluorescence Spectrometry

CHEN Liang,CHI Yan-Hua,SHANG Li-Ping,WANG Han
DOI: https://doi.org/10.3969/j.issn.1004-8138.2009.02.072
2009-01-01
Abstract:Fluorescence method was applied to the determination of anthracene and pyrene.The calibration curve linear ranges of anthracene and pyrene were 5.6×10-9—2.8×10-6mol/L and 4.9×10-8—1.47×10-6mol/L,respectively.Correlative coefficient was obtained as 0.99959 and 0.99822.The effect of common ion,temperature and time on the fluorescence intensity were studied.The no interference of common ions were for determination.Under the experiments conditions of 2 hours,5—55℃,the intensity of fluorescence was stable.Therefore this method has a wide range for temperature.can be used to detect anthracene and pyrene straight in non water system such as food,drink and liquor.
What problem does this paper attempt to address?