Enantiomeric Separation of Free Leucine by Pressurized Capillary Electrochromatography with Chiral Ligand Exchange Method

辛艳飞,万丽丽,马丽萍,孙烨,周向军,王永祥
DOI: https://doi.org/10.3969/j.issn.1004-4957.2008.10.015
2008-01-01
Abstract:A method for the separation of D-and L-leucine enantiomers by pressurized capillary electrochromatography with a chiral ligand exchange system using a chiral mobile phase additive and a reversed-phase C18 column was established.An acetic buffer(20 mmol/L,pH 5.5)containing 2 mmol/L of aspartame,sufficiently complexed copper Ⅱ ions and methanol,was used as the mobile phase.The operation pressure and voltage were 6.9×106 Pa and+5 kV,respectively.The eluent was monitored with a UV detector at 280 nm.The separation was completed in 15 min.The standard curves of D/L-Leu ranged linearly from 0.05 to 1.5 mmol/L(r 0.99).The within-day and between-day precisions of the quality control samples were less than 3.5% for both L-Leu and D-Leu.