Comparison of computed tomography angiography with colorduplex ultrasonography for measuring the maximal diameter of abdominal aortic aneurysm

LI Weimiao,CHEN Bin,GUO Daqiao,DONG Zhihui,ZHU Ting
2013-01-01
Abstract:Objective To assess the agreement of various colorduplex ultrasonography (CDU) measurementswith computed tomography angiography (CTA) for measuring the maximaldiameter of abdominal aortic aneurysm. Methods From April 2010 to June 2011, 77 patients were diagnosed as abdominal aortic aneurysm in our center. CDU and CTAwere applied in all patients before intervention. Maximal aorticdiameterwas recorded from intimal to intimal in the anteroposterior plane (DUAP_i), adventitia to adventitia in the anteroposterior plane (DUAP_a), transverse plane (DU_trans) and at anydirection (DU_max) by CDU. The CDUdata were compared with maximaldiameter obtained by CTA (CT_max). Correlation was assessed with linear regression analysis. Agreement between different measurements was evaluated according to the principles of Bland-Altman plots. Limits of agreement (LOA) between -5mm and 5mm were regarded as clinically acceptable. Results CT_maxwere significantly higher than DUAP_i, DUAP_a, DU_trans and DU_max (all P<0.05). The correlation coefficients between CTA and the four CDU methods (DUAP_i, DUAP_a, DU_trans and DU_max)were 0.968, 0.969, 0.947 and 0.990, respectively (all P<0.05). In Bland-Altman plots, the LOAs of the four pairs were -0.43 to 12.43, -3.80 to 8.70, -3.54 to 12.68 and -3.08 to 4.10 mm, respectively.0nly DU_max was clinically acceptable. Conclusion All CDU measurements correlate with CTA measurement, but only DU_max is in good agreement with CTA. Thus, replacing CTAwith CDU is feasible only when applying the DU_max methodology. (Shanghai Med J,2013,36:354-357)
What problem does this paper attempt to address?