An Analysis on False Negative Findings in Universal Newborn Hearing Screening

Tao Zheng,Wu Hao,Li Yun
DOI: https://doi.org/10.3969/j.issn.1006-7299.2007.01.006
2007-01-01
Abstract:Objective To study the validity of the use of otoacousitc emission in newborn hearing screening and the causes for false negative findings. It is important to monitor the hearing in infants at high risk.Methods Collected were the data of infants identified with hearing impairment in our hearing center from January,2002 to December,2005. 5 cases of newborns who passed the hearing screening were found to have hearing impairment. Their case histories, clinical examinations, audiometry and imaging testing results were carefully analyzed.Results Among the 5 patients, 2 cases were found with moderately severe sensory hearing loss and 3 cases with profound sensory hearing loss, including 1 case with auditory neuropathy. Auditory screening by OAE showed 99.38% sensitivity,0.62% false negative rate.Conclusion Auditory screening by OAE had high sensitivity, but also there were a few false negative cases in practices.Some cases with hearing loss due to retrocochlear disease would be missed.In addition,it's necessary to pay attention to delayed and hereditary sensorineural deafness.This should be given adequate attention in hearing screening.
What problem does this paper attempt to address?