A Voltammetric Sensor Based on Graphene-Modified Electrode for Simultaneous Determination of Catechol and Hydroquinone

Haijun Du,Jianshan Ye,Jiaqi Zhang,Xiaodan Huang,Chengzhong Yu
DOI: https://doi.org/10.1016/j.jelechem.2010.10.002
IF: 4.598
2011-01-01
Journal of Electroanalytical Chemistry
Abstract:In this report, a voltammetric sensor for simultaneous determination of hydroquinone (HQ) and catechol (CC) was developed at a glassy carbon electrode modified with graphene (GR/GCE). The separation of oxidation and reduction peak (ΔE) is decreased from 281 to 31mV for HQ and from 250 to 26mV for CC at GR/GCE, respectively. Separation of the oxidation peak potentials for HQ and CC was about 112mV in 0.10M acetate buffer solution (pH 4.5), and the anodic currents for the oxidation of both HQ and CC are greatly increased at GR/GCE, which makes it suitable for simultaneous determination of these compounds. Under the optimized condition, the anodic peak current of HQ is linear with the concentration of HQ from 1×10−6 to 5×10−5M in the presence of 5×10−5M CC. A detection limit of 1.5×10−8M (S/N=3) can be achieved. At the same time, the anodic current of CC is linear with the concentration of CC from 1×10−6 to 5×10−5M with a detection limit of 1.0×10−8M (S/N=3) in the presence of 5×10−5M HQ. The proposed sensor was successfully applied to the simultaneous determination of HQ and CC in tap water, and the results are satisfactory.