The Comparison of Two Prediction Models Based on Near-Infrared Three Wavelengths of Glucose Concentrations Measurement in Aqueous Glucose Solutions

WANG Long,XIAO Jun,LUO Qing-ming,GONG Hui
DOI: https://doi.org/10.3969/j.issn.1672-6278.2006.02.005
2006-01-01
Journal of Biomedical Engineering Research
Abstract:The modeling methods was discussed to determine glucose concentrations based on near-infrared three wavelengths in aqueous glucose solutions in the text.The model was established based on partial least-square algorithm.When it was used to predict the glucose contents of validation set,the root mean square error of prediction(RMSEP) equaled 17.08mg/dL and the correlation coefficient between prediction values and reference values attained 0.997;while using logarithmic ratio of absorbency difference method to set up prediction model,the RMSEP reached 54.94mg/dL and the correlation coefficient equaled 0.949 merely.The results suggest that the model based on PLS algorithm has lesser RMSEP,it also validates the system′s feasibility primarily.