Electrophysiology Detection and Electron Microscopy of Recurrent Laryngeal Nerve Injuries Induced by Ultrasonically Activated Shears

ZHOU Zheng,WANG Jia-dong,XIE Ming
DOI: https://doi.org/10.3969/j.issn.1674-8115.2007.09.007
2007-01-01
Abstract:Objective To determine the related factors of recurrent laryngeal nerve(RLN) injuries induced by ultrasonically activated shears(UAS) by electrophysiology detection and electron microscopy. Methods Fifteen New Zealand rabbits (thirty recurrent laryngeal nerves) were randomly divided into 5 groups(n=6): normal (control group), UAS continued working 2 s 1 mm beside RLN (group 1), 4 s 1 mm beside RLN(group 2), 2 s 3 mm beside RLN(group 3), 4 s 3 mm beside RLN(group 4). Electrophysiology of amplitude and latency of the injury were detected in each group. The injured RLN of group 2 and group 3 and RLN of the control group were observed with electron microscopy. Results There were significant differences in amplitude and latency between group 1, 2, 4 and control group, and there were significant differences among group 1, 2, 3 and 4. It was indicated by electron microscopy that in group 2 there were obvious myelin lamellar swelling, structure loose and disorder, axon collapse of a bubble and homogeneous slurry of axoplasm. The normal structure of RLN in group 3 and control group was observed. Conclusion UAS may lead to RLN injury. The severity of injury may increase with the shortening of distance between UAS and RLN and the extension of working time.
What problem does this paper attempt to address?