Fast Determination of Sugars in Coke and Diet Coke by Miniaturized Capillary Electrophoresis with Amperometric Detection

QC Chu,L Fu,YQ Guan,JN Ye
DOI: https://doi.org/10.1002/jssc.200400015
IF: 3.614
2005-01-01
Journal of Separation Science
Abstract:The fast separation capability of a novel miniaturized capillary electrophoresis with amperometric detection (CE-AD) system was demonstrated by determining sugar contents in Coke and diet Coke with an estimated separation efficiency of 60,000 TP/m. Factors influencing the separation and detection processes were examined and optimized. The end-capillary 300 microm Cu wire amperometric detector offers favorable signal-to-noise characteristics at a relatively low potential (+0.50 V vs. Ag/AgCl) for detecting sugars. Three sugars (sucrose, glucose, and fructose) have been separated within 330 s in a 8.5 cm length capillary at a separation voltage of 1000 V using a 50 mM NaOH running buffer (pH 12.7). Highly linear response is obtained for the above compounds over the range of 5.0 to 2.0 x 10(2) microg/mL with low detection limit, down to 0.8 microg/mL for glucose (S/N = 3). The injection-to-injection repeatability for analytes in peak current (RSD < 3.6%) and for migration times (RSD < 1.4%) was excellent. The new miniaturized CE-AD system should find a wide range of analytical applications involving assays of carbohydrates as an alternative to conventional CE and micro-CE.