Determination of Lycopene in Fruit and Vegetable Juice Drink by HPLC

Chen Xiang-Ming,Zhao Juan-Juan,Lv Qing-Zhi,Zhang Xiao-Fan,Sun Ji-Liang
DOI: https://doi.org/10.3969/j.issn.1004-8138.2012.04.067
2012-01-01
Abstract:Lycopene in fruit and vegetable juice drink was determined by high performance liquid chromatography(HPLC).Lycopene was extracted from drink by ultrasound extraction with acetone as solvent,and was separated by Ultimate-Cig column(250mm×4.6mm,5μm)with acetonitrile-ethyl acetate as the mobile phase at a flow rate of 1.0mL/min,and the detection wavelength was 475nm, and external standard calibration curve method was adopted to quantitate.There was the good linear relationship for lycopene in the range of 1.0×10~(-2)-3.9×10~(-5)g/L with correlation coefficient of 0.9999,and the detection limit was 6.3×10~(-6)g/L.The method has fast analysis speed,high sensitivity,wide linear range and good repeatability.