Simultaneous determination of inorganic anions and cations by ion chromatography with indirect photometric detection

Jinzhang Gao,Huijie Bian,Jingguo Hou,Jingwan Kang
DOI: https://doi.org/10.1016/0021-9673(93)83039-U
IF: 4.601
1993-01-01
Journal of Chromatography A
Abstract:Inorganic cations and anions were separated simultaneously on a silica-based Shim-pack WAX-1 column. The method is shown to provide efficient separation of Ca2+, Mg2+, Cl− and NO3−. Using 1 mM phthalate—6 mM EDTA at pH 6.57 as the eluent, a mixture containing these ions was separated in less than 6 min with detection limits of 1.2–2 ng. The effects of the eluent pH, concentration and composition on the retention behaviour of the analytes were studied. Two practical applications are described.
What problem does this paper attempt to address?