Study on sterilization of exhaust high efficiency particular air filter in situ in main area within biosafety level 3 laboratory

HAN Wen-Dong,SUN Zhi-Ping,DING Yue-Na,CAO Ming-Bo,QU Di
DOI: https://doi.org/10.3969/j.issn.1673-6184.2012.03.003
2012-01-01
Abstract:Laboratories—General Requirements for Biosafety(GB19489-2008) required "Exhaust high efficiency particulate air (HEPA) filter of biosafety level 3 laboratory should be sterilized in situ". Exhaust HEPA filter is an important defensive line to ensure the biosafety of BSL-3 laboratory facilities and also the main line of defense to ensure the non-leakage of infectious aerosols generated in the experimental procedure. In situ disinfection of the HEPA filter is needed in order to protect the safety of personnel. This study designed and produced a cycle disinfection prototype (with the function of formaldehyde volatilization, gas humidification and air circulation supply). The prototype was connected with the disinfection interface of HEPA filter’s exhaust box. And then the steam cycle of formaldehyde was driven through the HEPA filter by air circulation supply system. Thus the in situ disinfection was realized. The results showed that formaldehyde vapor concentration at room temperature was 16 mg/L, and the relative humidity was 70%. After 12-hour continuous formaldehyde fumigation, Staphylococcus aureus, Mycobacterium smegmatis, Bacillus subtilis var. niger and Bacillus stearothermophilus were completely disinfected. As a consequence, the in situ disinfection of HEPA in both equipment and methodology is achieved.