Development of an assay for the detection of the survival rates of cells by flow cytometry

Liu Yang,Lai Yi,Li Minhui,Yang Shuxia,Zheng Jing,Yang Tai,Li Limei,Deng Yong,Zhang Qiang,Ren Rongmei,Li Qianmei,Zhang Shuai,Zou Qiang,Li Hua
DOI: https://doi.org/10.3969/j.issn.1673-4130.2011.15.003
2011-01-01
Abstract:Objective To investigate the feasibility and accuracy of flow cytometry(FCM) for the detection of the survival rates of cells to provide another choice of drug screen,cell proliferation assay and drug toxicity assay.Methods Flow cytometry,CCK-8 and trypan blue staining technique was performed for the detection of the survival rates of Jurkat cells.Results The coefficient of variation(CV) of FCM was less than 5%,when the cell concentration exceed 105/mL and was 5%-10%,when the cell concentration was below 105/mL.The r2 and slope of FCM were 0.993 5 and 0.999 4 respectively,those of CCK-8 were 0.991 2 and 1.011 respectively and those of trypan blue staining technique were 0.969 8 and 1.057 respectively.Conclusion The flow rate of sample introduction of FCM could be stable and with direct proportion with the amount of viable cells.FCM could be used for the detection of the survival rate of cells.
What problem does this paper attempt to address?