Capillary Electrophoretic Method for Determination of Matrine in Caco‐2 Cell Medium

PCH Li,GH Gao,FCP Law
DOI: https://doi.org/10.1081/jlc-200030498
2004-01-01
Abstract:A capillary electrophoretic (CE) method has been developed for the analysis of matrine in a biological medium. The optimized run buffer solution contained 200 mM tris(hydroxymethyl)aminomethane, 40 mM sodium dihydrogen phosphate, and 20% 2-propanol, adjusted to pH 5.8 using phosphoric acid. A voltage of 25 kV was applied across a capillary (50 mum x 40 cm, distance to detector: 10.2 cm) for CE separation. Each analysis was completed in less than 5 min. Linearity of matrine calibration in the concentration range of 3.2-185 mug/mL (at the detection wavelength of 214 nm) was excellent (R-2=0.9994). The run-to-run repeatability (n=3), as expressed by the relative standard deviation (RSD), was found to be better than 4.0%. The limit of detection was estimated to be 3.2 mug/mL. This CE method was employed to determine matrine in Caco-2 cell media after transport of matrine through the cells.