Quantitative Analysis of Glyphosate by Infrared Absorption Spectrometry

DAI Yu-jing,YAO Jie,FENG Yu-ying
DOI: https://doi.org/10.3969/j.issn.1006-0413.2011.10.010
2011-01-01
Abstract:[Methods] The content of glyphosate was measured by infrared spectroscopy with ATR.Absorption ranged from 4000-500 cm-1.Resolution was 1 cm-1 and scan time was 32.[Results] A comparison of infrared spectra between standard glyphosate and sample glyphosate showed that the absorption peak of glyphosate at 1321 cm-1 was independent and could be used as the characteristic absorption peak for quantitative determination.The area from 1295 to 1348 cm-1 of the infrared spectra of glyphosate and its content met the linear relationship with r 0.999 7.[Conclusions] The method of ATR-FTIR,instead of routine lab analysis,is feasible for the determination of glyphosate in agrochemical.