Determination of related substances and content of oroxylin A by RP-HPLC
Jia-zhuo LIU,Ruo-qi WANG,Wen-yuan LIU
DOI: https://doi.org/10.7539/j,issn.1672-2981.2019.01.021
2019-01-01
Abstract:Objective To establish an RP-HPLC method to determine the related substances and content of oroxylin A. Methods RP-HPLC was adopted. ODS-2 C18 column (4.6 mm×150 mm, 5 μm) was used with a gradient elution system of acetonitrile-methanol-10 mmol·L-1 phosphate buffer. The column temperature was 30℃, the flow rate was 1.0 mL·min-1, and the detection wavelength was 271 nm. Results A linearity between the area of oroxylin A and its concentration was found at 2.02-20.23 μg·mL-1 (r = 1, n = 6). For the determination of impurities, the standard curves of baicalein, wogonin and impurity Ⅰ were linear at 0.0197-0.7872 μg·mL-1 (r = 1.0000, n = 6), 0.0313-0.6250 μg·mL-1 (r = 0.9997, n = 6), and 0.0818-2.0440 μg·mL-1 (r = 0.9999, n = 5), respectively. The contents of baicalein, impurity Ⅰ and the total content of the impurities were 0.11%, 0.39%, and 0.57%, respectively. Conclusion The method is sensitive, reproducible, accurate and reliable in the determination of related substances and quality control of oroxylin A in the raw material.